Wynne Corrector Project
under the direction of Dr. Edwin Loh
Michigan State University
REU-summer 1997
___________________________________________________________________________________________
Abstract
This paper includes the design and evaluation of the current Wynne corrector system for a 2.3-meter telescope. It
also summarizes the specification for, design, and evaluation of the new Wynne corrector that I worked on this
summer.
_________________________________________________________________________________________
History
The University of Wyoming maintains a telescope observatory near Laramie, Wyoming. At this observatory
stands an f / 2 telescope that is 2.3 meters in diameter. Professor Edwin Loh of Michigan State University
designed the original Wynne corrector lenses for that telescope. The old corrector system was used to obtain a
field of 15 arcmin. My project was to design a series of corrector lenses for a 1.2-degree field, which is 5 times as
large.
Wynne (1987) Correctors are a grouping of either three or four lenses in a telescope that correct for the
aberrations of the primary mirror.
1.0 Original Wynne Corrector
My REU project was to help design a new Wynne corrector for the telescope in Wyoming by modifying the
design of the original corrector. The following information was obtained using an optics program called Zemax and
notes from Dr. Edwin Loh at Michigan State University.
Some specifics for the Wynne corrector and telescope currently in place:
It is an f / 2 telescope.
The entrance pupil diameter of the telescope is 2,300 mm.
The focal length from Zemax is 4,780.77 mm. (This should theoretically be 4,600 mm but the corrector increases the focal length slightly.)
BK7 glass was used for all surfaces except the mirror.

Figure 1: The above graph shows the 4.6-meter
telescope with the original Wynne corrector lenses.
(All dimensions are scaled to best depict the actual system.)

Figure 2: The above diagram shows a closer view of the
original Wynne corrector lenses for the 4.6 meter
Wyoming telescope.
1.1 Preliminary Testing of System
Using the above setup produced very good results. I used values of 0.8, 0.55, and 0.44 microns for the
wavelengths to test the spot diagrams of the system. I also sent bundles of rays into the system at angles of 0, 0.1,
and 0.12 degrees. The value of 0.12 degrees is very near the outer range for the system. By using that outer value,
I could see how the system reacted to degree values that were very near its limitations. I received spot diagrams
for the three different degree values with root-mean-square sizes of 2.5, 5.2, and 6.4 microns respectively. This
was much improved over the values that were taken with just the mirror, which has coma. The values obtained
with just the mirror were 0.3, 190.1, and 228.2 microns.
1.2 Surface Data Summary of Original Telescope System
The following chart shows the RMS values for the three different fields at each of the four wavelengths.
The wavelengths and RMS values are given in microns, and the fields are given in degrees.
|
|
0.00 |
0.10 |
0.12 |
|
0.34 |
6.360 |
8.074 |
9.085 |
|
0.44 |
3.880 |
5.498 |
6.561 |
|
0.55 |
2.644 |
4.526 |
5.703 |
|
0.80 |
1.619 |
4.288 |
5.608 |
|
ALL |
4.035 |
7.519 |
8.544 |
Following is the lens data for the original four-lens corrector system.
|
Surface # |
Radius Of Curvature |
Thickness |
Glass Type |
Position |
Diameter |
||||
|
1 2 3 4 5 6 7 8 9 10 11 12 13 |
-9200.00 -85.36 -94.31 -287.10 -132.70 -82.12 -50.42 -144.20 389.60 INFINITY INFINITY INFINITY INFINITY |
4606.88 11 35 9 50 6 25 8 35 3 10 3 4.88 |
Air BK7 Air BK7 Air BK7 Air BK7 Air BK7 Air BK7 Air |
4606.88 199.88 188.88 153.88 144.88 94.88 88.88 63.88 55.88 20.88 17.88 7.88 4.88 |
2300 110 110 90 90 64 64 60 60 40 40 40 40 |
||||
|
(All units are in millimeters.) |
|||||||||
1.3 Tolerance Testing of System
I checked the tolerance of the system to errors in positioning the lenses with respect to the mirror. I studied the
changes in the RMS radius of the spot diagrams while I varied the decentering of the lenses in the x and y
direction. I also evaluated how it would effect the system if the lenses were tilted in the x or y direction. I found
that the lenses could only be off center approximately 0.15 mm for there to still be a reasonable sized spot
diagram. I found that decentering by a value in the x direction makes the spot diagram larger than by decentering
by that same value in the y direction. I found that the lenses could be tilted about the respective axis by an angle of
approximately 0.07 degrees before the spot diagrams became too large by our standards.
1.4 Corner Images
I evaluated what happens to the images at the corners of the pictures when the system is tilted or decentered. I
changed the field location to 0.175 degrees. At this angle, the fields are outside of the circle of good images, but
still in the square of the detector. I also checked to see what happens to the corners when the length between the
mirror and the other lenses is varied. The changes in the corner images for decentering and tilting were hard to
distinguish in the range of change that I was working with. The moving back and forth of the mirror with respect to
the other lenses made a large difference though. When I made the distance between the mirror and the lenses
smaller, the image took on an almost "fan-like" pattern. One end was narrow and the other feathered out like a
fan. When I increased the distance between the mirror and the lenses from 4407 mm, the image spread out into
almost a triangle or diamond shape. It was very easy to tell from the test images if the distance between the
corrector and the mirror is too long, too short, or the correct length.

(System too short by 6 mm)

(Correct system length: 4,407 mm)

(System too long by 6 mm)
The above pictures are examples of what the spot diagrams from the original system looked like. These were without decentering or tilting of the system. I was just testing here what effect changes in the length of the system would make on the spot diagrams.
2.0 New Four-Lens Wynne Corrector
2.1 Preliminary Design
I designed the new Wynne corrector. I started by making the corrector lenses 4 times bigger. Almost every value
from the original system was multiplied by four and put into Zemax for evaluation of the new system. The focal
length of the telescope was kept the same because the new lens system must replace the old one directly in the
same telescope when it is completed. I also changed the weight function that acts on the different degree fields so
that it put more emphasis on the spot diagrams toward the outer range of approximately 0.6 degrees. I did that
because the program was trying to really focus in on the center regardless of how large the outer-range spot
diagrams were becoming.
There were many different variables I had to take into consideration while designing this new system. I allowed the
Zemax program to change the radii of curvature for the four corrector lenses in order to optimize the system. I
also let the program change the distances between the corrector lenses. Zemax has an "optimization" function that
finds the placement of the system that gives the smallest RMS values possible. I soon realized that Zemax was
trying to allow the edge thickness of the lenses to be larger than we wished them to be. It would be in our best
interests to have the lenses as thin and narrow as physically possible so as to minimize the cost of producing the
new lenses.
The original Wynne corrector lenses were made of BK7 glass. There were four different glass types that I
considered while designing the new system. The glass types were BK7, F2, F6, and LaSFN9. I chose to analyze
the other three glass types (besides BK7) because their indexes of refraction are very close to that of BK7.
Therefore, all four glass types are somewhat similar in optical nature.
After my evaluation of the system with each of the four types I found that BK7 was still the best glass to use. The
RMS values produced by the other three types of glass were greater than that of BK7 in all of my computations.
2.3 The Best Four-Lens Design
The following chart shows the RMS values for the five different fields at each of the four wavelengths.
The wavelengths and RMS values are given in microns, and the fields are given in degrees.
|
|
0.00 |
0.15 |
0.30 |
0.45 |
0.60 |
|
0.34 |
10.562 |
10.762 |
12.199 |
17.074 |
27.519 |
|
0.44 |
4.336 |
4.309 |
3.837 |
4.202 |
10.972 |
|
0.55 |
1.792 |
4.064 |
6.311 |
6.525 |
6.871 |
|
0.80 |
7.550 |
9.753 |
13.358 |
15.240 |
14.235 |
|
ALL |
6.902 |
8.043 |
10.143 |
11.849 |
14.431 |
Following is the lens data for the four-lens corrector system that produced the smallest RMS values.
|
Surface # |
Radius Of Curvature |
Thickness |
Glass Type |
Position |
Diameter |
||||
|
1 2 3 4 5 6 7 8 9 10 11 12 13 |
-9200.00 -401.8211 -451.8972 -551.7484 -366.602 -443.7244 -219.2816 -280.4669 11043.4668 INFINITY INFINITY INFINITY INFINITY |
3771.27 44 196.003 36 214.5867 24 204.6393 32 31.38873 12 40 12 20.99656 |
Air BK7 Air BK7 Air BK7 Air BK7 Air BK7 Air BK7 Air |
4639.8849 867.61429 823.61429 627.61129 591.61129 377.02459 353.02459 148.38529 116.38529 84.99656 72.99656 32.99656 20.99656 |
2300 440 440 360 360 256 256 240 240 160 160 160 160 |
||||
|
(All units are in millimeters.) |
|||||||||
3.0 Trying a Three Lens System
I designed two different three-lens corrector systems. Making a system that requires only three lenses would
greatly diminish the costs of the finished product. Is a three-lens system's effectiveness comparable to a four-lens
system? That was the next question that I had to answer.
Wynne (1974) described a three-lens corrector system.. His specifications were for an f / 3.25 telescope. I
modified this corrector lens package for an f / 2 telescope. I gradually changed the length of the system until it was
f / 2. After that, I slowly modified the thickness of the lenses until they were very near the values that I had been
testing on the four-lens system. The Wynne design that I modified from the journal article did not work better than
the four-lens system. In fact, it was much worse than the four-lens system that I had designed. The RMS values of
the three lenses were 10.2, 13.6, 21.3, 28.2, and 36.2 microns. These range from approximately a factor of two
to four times worse than the values of 5.3, 6.5, 8.5, 9.1, and 9.5 that I received from the four-lens system.
I experimented with a three-lens system of my own. I took the four-lens system that I already had and just deleted
out the smallest lens in the system. I then optimized the system and randomly changed a few of the distances and
semi-diameters just to see what would happen. Somehow, I came up with a very efficient system that gave me
RMS values of 7.1, 7.3, 8.4, 10.6, and 13.4 microns. Those values are very good when compared to the values
that I had been receiving from the system that I modified from the Wynne (1974) design. Also, the numbers are
fairly close to the values obtained from the four-lens system.
The following chart shows the RMS values for the five different fields at each of the four wavelengths.
The wavelengths and RMS values are given in microns, and the fields are given in degrees.
|
|
0.00 |
0.15 |
0.30 |
0.45 |
0.60 |
|
0.34 |
17.681 |
18.002 |
19.577 |
23.402 |
29.458 |
|
0.44 |
6.784 |
8.073 |
11.049 |
14.423 |
17.301 |
|
0.55 |
3.298 |
3.204 |
4.126 |
6.910 |
10.650 |
|
0.80 |
9.946 |
9.477 |
8.851 |
10.016 |
13.792 |
|
ALL |
10.822 |
11.254 |
12.735 |
15.300 |
18.423 |
The following chart contains the values for the three-lens corrector system that I prepared.
|
Surface # |
Radius Of Curvature |
Thickness |
Glass Type |
Position |
Diameter |
|
1 2 3 4 5 6 7 8 9 10 11 |
-9200 -385.6787 -425.4223 -716.7905 -243.2438 -358.4469 2696.547 INFINITY INFINITY INFINITY INFINITY |
3731.79 35.3661 373.542 24.3638 329.8486 18.49218 51.85736 12 40 12 27.06344 |
Air BK7 Air BK7 Air BK7 Air BK7 Air BK7 Air |
4656.32348 924.53348 889.16738 515.62538 491.26158 161.41298 142.9208 91.06344 79.06344 39.06344 27.06344 |
2300 400 400 210 210 190 190 160 160 160 160 |
(All units are in millimeters.)
References:
This report was prepared by Colleen Lurz on August 7th, 1997 at Michigan State University
for a Research Experience for Undergraduates (REU) program that was
sponsored by the National Science Foundation.